COLD
JUNCTION OFFSET CALIBRATION (C.J.OF)
The cold junction offset equals the actual temperature minus the reading
temperature. You may compensate for any error due to cold junction on
the TC input. You may perform this compensation in any temperature from
0░C to 40░C (or 32░F to 104░F), however we recommend you perform this
compensation at 0░C (32░F) for best result. To do this, immerse the thermocouple
hot junctions into a mixture of ice and water. Check the Reading Configuration
bit R.1 of the "RD.CF" menu setting for the proper temperature units.
TO PERFORM THIS COMPENSATION, FOLLOW THESE STEPS:
1. Connect the thermocouple wire to the +S and -S input.
2. Press the MENU button until meter displays "CJ.OF".
3. Press the /DEV
button. the meter displays the previous offset value with flashing 4th
digit.
4. Press the /DEV
button again. The reading temperature will be displayed (with no digit
flashing).
5A. If the value is okay, then press the MENU button.
The display will show "STRD" and 0 value will be entered at the
offset.
B. If the value is not okay, then enter the actual temperature
using the /DEV
and /MAX
buttons. Once you enter the accurate temperature, press the MENU
button. The meter displays "STRD" and stores the offset value.
1.
Temperature unit is either celsius or fahrenheit and will always be displayed
at 0.1 degree resolution. The meter flashes corresponding LED.
2.
MAX/MIN offset value will be ▒25.0 ░C or 45.0 ░F. If offset the limit,
the meter will flash "ER 3" and previous offset will not be changed.
LOCK
OUT CONFIGURATION ("LK.CF")
Use Lock Out Configuration
("LK.CF") for the following:
* To enable or disable setpoint changes
* To enable or disable the RESET button in the run mode
To
Enable or Disable the RESET button in the Run Mode
1. Press the MENU button until the meter shows "LK.CF"
after ("C.J.OF").
2. Press the DEV
button until the meter shows "RS.=E" (Default).
3. Press
the /MAX button
to toggle between:
"RS.=E" To
enable the RESET button in the run mode (default)
"RS.=D" To disable the RESET button in the run mode.
4. Once desired
mode shows, press the MENU button to store the change. The meter
returns to the run mode.
To
Enable or Disable Setpoint Changes
1. Press the MENU button until the meter shows "LK.CF"
(after "C.J.OF").
2. Press the /DEV
button. The meter shows "SP.=E" (Default).
3. Press the /MAX
button to toggle between:
"SP.=E" To enable setpoint changes (default)
"SP.=D" To disable setpoint changes
4. Once
desired mode shows, Press the MENU button to store the change.
To
Enable DisplayÆs Program Version:
1. Press
the MENU button until the meter shows "LK.CF" (after "C.J.OF").
2. Press
the /DEV
button. The meter shows one of the following:
* "L.3=0" ôSETPTSö button will display setpoint values.
* "L.3=1" "SETPTS" button will display "V.-03" which is the meter's
current software version.
2. Press the /MAX
button to toggle between the choices above.
3. Press the MENU button to store the changes.
If your meter does
not have the relay option, setpoint menu items above will not be available
and SETPTS button will always display the meter's software version.
These units will have +OL (overload) or +OPEN memory indicated
by Alarm 1 & 2 LED displays. LEDs can be reset by pressing MENU then
RESET button or by power OFF then ON. These units
can not use analog output proportional to error from setpoint 1. under
menu OT.CF, 0.3=P
TUNING
PROPORTIONAL CONTROLLER
The Proportional Controller is not active unless your
meter has analog output capabilities. The menu will display whether analog
output is present or not. This function allows you to tune your controller
provided you have analog output capabilities. Select proportional on Output
Configuration (refer to Section 15-3) prior to tuning your controller.
Include the meter in the process loop and turn on the meter. Allow enough
time for the system to settle, then do the following.
1. Press the /DEV
button. The meter momentarily shows "DEV" followed by a blinking
value. This value is the deviation (error) between Reading and Setpoint
1 values. If this error is zero, your controller is tuned. If a value
other than zero shows, proceed with step 2.
2.. Press RESET button. The meter shows "TUNE", tuning
your controller and canceling any error. Once tuned, the meter shows "RST"
and returns to the run mode.
Allow
enough time for process to settle before proceeding with "TUNE"
procedure. If any error happens during this procedure, meter will flash
"ER 4" and abort the tuning. You have to restart the procedure.
"TUNE" will be active if your meter has analog output
capabilities.
3.
Press
the /DEV
button. Verify that
blinking value is zero.
If blinking value is
not zero, repeat step
2. |